exocad Bite Splint module: occlusal splint design step by step

How to design an occlusal splint in the exocad Bite Splint module: scans and bite, border and thickness, contacts and guidance, then printing or milling it.

On a render an occlusal splint looks like a simple plate of resin over the teeth. What decides whether it works is invisible there: where the contacts are, how the guidance behaves in movement and whether the splint seats without rocking. In exocad the design runs in the Bite Splint module, ideally together with the Virtual Articulator.

The splint type and the therapeutic jaw position are clinical decisions. This guide starts from the point where both are defined.

Scans and records for an occlusal splint

The design needs scans of both arches, with the teeth and every area the splint will cover fully captured. The bite registration must be taken in the position the clinician wants for the splint, usually with a defined vertical opening. Articulator values or jaw motion data make the dynamic contacts reliable (see virtual articulator and jaw motion).

The order should state the arch, the splint type, the desired thickness, the guidance concept and the material.

Designing the splint in the Bite Splint module

Insertion direction and undercuts

Define the insertion axis and how much undercut the splint may use for retention. With too little undercut the splint drops out. With too much it does not seat, or it hurts to put in.

Splint border

Draw the border on the teeth and, where needed, on the soft tissue. A consistent border height retains the splint without pressing on the gingiva.

Splint body

The module builds the body against the opposing arch. Set the minimum thickness from the material data sheet and from the clinician’s request.

Occlusal surface and contacts

Contacts with the antagonist are created on the occlusal surface. The surface can then be smoothed and shaped and exocad can flatten the posterior occlusal area automatically.

Guidance and interferences

In the Virtual Articulator, run the movements, build canine guidance where it was requested and remove the interferences. Each adjustment removes material, so check the thickness again afterwards.

Final checks and export

Add text or a logo if the clinic wants one, look at the final thickness map and export a closed mesh for printing or milling.

Printing or milling the splint

Printing Milling
Speed Several splints per build One blank per splint
Material cost Low Higher
Material Resins indicated for splints Pre-polymerized blanks
What to control Validated printer, resin and post-curing Tool strategy and blank size

Both routes work when the material is indicated for splints. Printed splints must follow the resin manufacturer’s validated parameters and post-processing exactly, because under-cured or poorly washed parts change both fit and material properties (see resin exposure calibration).

Why splints need adjusting chairside

When a splint needs a lot of grinding at delivery, the cause is usually in the records or in a skipped check. A bite registration with the wrong vertical opening shifts every contact. Contacts designed only in static occlusion interfere as soon as the patient moves. Borders set too high or undercuts too deep make the splint hard to seat. Adjustments made after the last thickness check can leave areas below the material minimum.

Frequently asked questions

Which scans are needed for an occlusal splint?

Both arches and a bite registration in the position the clinician wants for the splint, usually with a defined vertical opening. Articulator values or jaw motion data make the dynamic contacts more reliable.

Should a splint be printed or milled?

Both work when the material is indicated for splints. Printing is fast and economical and milling from pre-polymerized blanks gives a very homogeneous material. Follow the material manufacturer's indications.

Can the Bite Splint module design different splint types?

Yes. Relaxation splints, night guards, repositioning appliances and deprogrammers can be designed. The therapeutic choice belongs to the clinician.