MDR and implant-compatible design files for dental labs

MDR and implant-compatible design files: what a dental lab needs before scaling custom abutments and bars, from prescriptions to Ti-base limits and records.

Implant work is where a dental lab grows fastest and where the regulatory questions get sharper. A custom abutment or a screw-retained bar combines a design from the lab, a component from another manufacturer and an implant from a third. This post explains what the EU Medical Device Regulation (MDR) and the MDCG 2021-3 guidance mean for MDR implant-compatible design work and what a lab should have in order before it scales up. It is general information, not legal advice: the competent authority and a regulatory consultant have the final word for a specific lab.

When implant-borne work is a custom-made device

The MDR defines a custom-made device as one made on the written prescription of an authorized person, with specific design characteristics given under that person’s responsibility, for the sole use of one patient. MDCG 2021-3, the official Q&A on custom-made devices, uses a dental crown made on a dentist’s prescription as its example.

The guidance also says what a prescription needs at minimum: the patient’s name, or a pseudonym where relevant, and design characteristics specific to that patient. Models, 3D model data and impressions can be part of those characteristics. Dimensions on their own, such as DICOM files from a CT scan, are not enough. For implant cases that means the order from the clinic should carry the clinical choices that only the dentist can make, such as the restoration type, the screw-retained or cemented decision and the material, together with the scans.

Two kinds of devices are excluded. Devices mass-produced by industrial processes are not custom-made, even when they follow a prescription. Neither are patient-matched devices, which are fitted to a patient within a design envelope and typically made in batches through a validated process under the sole responsibility of a manufacturer. Those follow the standard MDR route for placing devices on the market.

Ti-bases, libraries and the components you do not make

Most implant-borne restorations from a lab are built on a prefabricated part: a Ti-base, a multi-unit abutment or a screw. MDCG 2021-3 confirms that parts and components intended for custom-made devices may themselves be CE-marked medical devices and that their instructions for use must be followed when they are prepared, assembled or adapted.

For the design that has concrete consequences:

  • the Ti-base’s instructions set limits such as the minimum wall thickness of the ceramic part, the maximum angulation and the gingival height and the design stays inside them;
  • the implant library in the CAD has to match the component actually used, from the supplier of that component, in a current version (see installing implant libraries in exocad);
  • the combination of implant, Ti-base and screw has to be one the component manufacturers state as compatible;
  • the lab records which component and batch went into each case.

A design that ignores a Ti-base limit, or a library that does not match the part in the box, moves the risk onto the lab as manufacturer of the final device. The differences between abutment types are covered in implant abutment types.

Who is the manufacturer when design is outsourced

When a lab sends the design to an outside designer and the milling to a production center, the device still has one manufacturer: normally the lab that receives the prescription and delivers the finished device to the clinic with the Annex XIII statement. The designer and the milling center are suppliers. Their work goes into the lab’s technical documentation and the lab chooses and controls them as part of its quality system.

In practice the lab keeps the design file with the case, records who designed and who produced it and agrees with each supplier on what they check and what they hand back. Outsourcing dental CAD design describes how a working arrangement looks.

Obligations that grow with volume

MDCG 2021-3 states that custom-made device manufacturers have to meet nearly all MDR requirements. A small lab can run them in proportion to its size, but they grow with the number and type of devices:

  • a quality management system proportionate to the risk class and the type of device, covering what Article 10(9) lists;
  • risk management, post-market surveillance and clinical evaluation, which the guidance allows to be run per group of devices with the same purpose, materials, process and principal design, rather than per single device;
  • a person responsible for regulatory compliance;
  • reporting of serious incidents and field safety corrective actions to the competent authority;
  • a post-market surveillance report for Class I devices, or a periodic safety update report for higher classes.

Custom-made devices do not carry UDI and the responsible person does not have to be registered in EUDAMED. The one case that needs a notified body is a Class III implantable custom-made device, which requires a certified quality management system under Article 52(8).

The risk class of each device type is set by the rules in Annex VIII of the MDR and implant-borne parts are worth checking one by one. A lab that adds custom abutments or bars to its crowns and dentures should review its classification and documentation for those devices before the volume grows.

Signs a lab is moving beyond custom-made

Using CAD/CAM, milling and printing does not change the status: MDCG 2021-3 states that modern technologies may be used for custom-made devices as long as the definition is met and the device is not mass-produced. Questions arise when a lab starts to make standardized parts in batches, sells designs or components to other labs as products, or produces without an individual prescription. At that point the regulatory route needs a fresh look with a consultant.

Sources

Frequently asked questions

Is a custom abutment designed in the lab a custom-made device?

It can be, when it is made for one patient on a written prescription with patient-specific design characteristics and is not mass-produced. The MDCG 2021-3 guidance gives a dental crown made on a dentist's prescription as a typical custom-made device and the same test applies to implant-borne work.

Can a lab use a CE-marked Ti-base from another manufacturer in its abutments?

Yes. MDCG 2021-3 confirms that parts and components intended for custom-made devices may be CE-marked medical devices. Their instructions for use have to be followed when the lab adapts or assembles them, which in practice means respecting the design limits stated for that Ti-base.

Does scaling up production change the status of custom-made devices?

Modern methods such as CAD/CAM and 3D printing are allowed for custom-made devices as long as each device meets the definition and is not mass-produced. A lab that starts producing standardized parts in batches, without individual prescriptions, is moving toward a different regulatory route.